Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). B3: DATE OF EVENT UNKNOWN / NOT PROVIDED. E1: LAST/GIVEN NAME UNKNOWN / NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER ¿ K241032. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMVIE FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. D9: DEVICE AVAILABILITY AND RETURN DATE WERE UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE ADDED: 4111 AND 3331. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 3252. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 4307. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE RECEIVED THE REPORTED BAR FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED. THE RETURNED BAR/PROSTHESIS WAS FRACTURED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. A LOT SPECIFIC COMPLAINT HISTORY REVIEW IS NOT CONDUCTED FOR PSP COMPLAINTS. PATIENT SPECIFIC PRODUCTS HAVE A UNIQUE ONE TIME USE LOT NUMBER. THEREFORE, A LOT SPECIFIC HISTORY SEARCH FOR THE BAR WAS NOT PERFORMED. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION HAS OCCURRED. THE RETURNED BAR WAS OBSERVED TO BE FRACTURED. THE REPORTED EVENT HAS BEEN CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
IT WAS REPORTED THAT THE BAR BROKE IN PATIENT'S MOUTH. WILL BE RETURNING THE BROKEN ONE AND REMAKING A NEW BAR. NO ADDITIONAL INFORMATION WAS PROVIDED.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1745561 | BELLATEK HYBRID BAR 4 IMPLANTS | DENTAL IMPLANT BAR | NHA | BIOMET 3I | 8724950-1 | 00844868001676 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Male |