FDA Adverse Event
Malfunction
Summary report: N
8800
MDR report key: 2102089
·
Received May 18, 2011
Report
- Report Number
- 9617766-2011-01084
- Event Type
- Malfunction
- Date Received
- May 18, 2011
- Date of Event
- May 3, 2011
- Report Date
- May 18, 2011
- Manufacturer
- GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE MAIN SOCKET AND PLUG WERE REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 8800 | FLUOROSCOPIC X-RAY | JAA | GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD | 8800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |