FDA Adverse Event Malfunction Summary report: N

MICRO DRILL LONG STRAIGHT ATTACHMENT

MDR report key: 2101990 · Received May 18, 2011

Report

Report Number
1811755-2011-01797
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 19, 2011
Report Date
April 19, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DURING FAILURE ANALYSIS, THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE DEVICE OVERHEATED DURING PERFORMANCE TESTING. VISUAL EXAMINATION REVEALED SEVERE CORROSION WITHIN THE INTERNAL COMPONENTS OF THE DEVICE. CORROSION DAMAGE AND DEBRIS WITHIN THE BALL BEARINGS WAS IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE PRESENCE OF CORROSION AND DEBRIS WITHIN THE COMPONENTS LEADS TO INCREASE FRICTION BETWEEN THE MOVING PARTS; THIS CAN RESULT IN INCREASED HEAT PRODUCTION AS REPORTED BY THE CUSTOMER. CORROSION DAMAGE IS KNOWN TO OCCUR OVER TIME DUE TO IMPROPER STERILIZATION METHODS AND REPEATED AUTOCLAVE CYCLES. THE STRYKER MICRO DRILL LONG STRAIGHT ATTACHMENT IS NOT A REPAIRABLE DEVICE.

Description of Event or Problem · 1

THE STRYKER MICRO DRILL LONG STRAIGHT ATTACHMENT WAS SENT IN FOR REPAIR AND IT OVERHEATED DURING QUALITY INVESTIGATION. NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO DRILL LONG STRAIGHT ATTACHMENT DZJ STRYKER INSTRUMENTS KALAMAZOO 10113

Patients

Seq Age Sex Outcome Treatment
1 UNK