FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2101988 · Received May 18, 2011

Report

Report Number
1720753-2011-07410
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
May 3, 2011
Report Date
May 18, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP CONDUCTED AN ONSITE INVESTIGATION. THE CANDLE STICKS WERE REGREASED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM MADE A LOUD POPPING NOISE AND THEN DISPLAYED A COMMUNICATIONS FAILURE ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1