FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2101958 · Received May 18, 2011

Report

Report Number
1811755-2011-01792
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 19, 2011
Report Date
April 19, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DURING FAILURE ANALYSIS THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE DEVICE OVERHEATED DURING PERFORMANCE TESTING. VISUAL EXAMINATION REVEALED WORN/DAMAGED MOTOR, ROTOR, DRIVE SHAFT AND SPINDLE HOUSING. ROTOR ROB, DAMAGED MOTOR CARTRIDGE AND SPINDLE HOUSING WERE IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE ROTOR COUPLER TRANSMITS TORQUE TO THE DRIVE SHAFT, AS A RESULT, ANY FAILURE OF THE COUPLER CAN CAUSE THE DRIVE SHAFT AND THE ROTOR TO STICK. THIS CAN LEAD AN INCREASE IN FRICTION BETWEEN THE MOVING COMPONENTS THUS RESULTING IN INCREASED HEAT PRODUCTION AS DESCRIBED BY THE CUSTOMER. THE DAMAGED PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO CIRCULATION.

Description of Event or Problem · 1

THE STRYKER CORE IMPACTION DRILL WAS SENT IN FOR EVAL DUE TO OVERHEATING DURING A PROCEDURE. NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE DEVICE; THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE WITHOUT ANY DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK