CORE IMPACTION DRILL
Report
- Report Number
- 1811755-2011-01792
- Event Type
- Malfunction
- Date Received
- May 18, 2011
- Date of Event
- April 19, 2011
- Report Date
- April 19, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- UNKNOWN
Narratives
DURING FAILURE ANALYSIS THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE DEVICE OVERHEATED DURING PERFORMANCE TESTING. VISUAL EXAMINATION REVEALED WORN/DAMAGED MOTOR, ROTOR, DRIVE SHAFT AND SPINDLE HOUSING. ROTOR ROB, DAMAGED MOTOR CARTRIDGE AND SPINDLE HOUSING WERE IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE ROTOR COUPLER TRANSMITS TORQUE TO THE DRIVE SHAFT, AS A RESULT, ANY FAILURE OF THE COUPLER CAN CAUSE THE DRIVE SHAFT AND THE ROTOR TO STICK. THIS CAN LEAD AN INCREASE IN FRICTION BETWEEN THE MOVING COMPONENTS THUS RESULTING IN INCREASED HEAT PRODUCTION AS DESCRIBED BY THE CUSTOMER. THE DAMAGED PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO CIRCULATION.
THE STRYKER CORE IMPACTION DRILL WAS SENT IN FOR EVAL DUE TO OVERHEATING DURING A PROCEDURE. NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE DEVICE; THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE WITHOUT ANY DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |