FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE 2 DOCKING STATION (120V)
MDR report key: 2101945
·
Received May 18, 2011
Report
- Report Number
- 1811755-2011-01809
- Event Type
- Malfunction
- Date Received
- May 18, 2011
- Date of Event
- April 18, 2011
- Report Date
- April 18, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SVC REP EVALUATED THE DOCKER AND FOUND THAT THE FAUCET LOCATED ABOVE THE DOCKER WAS CORRODED AND LEAKING WATER DIRECTLY ONTO THE DOCKER. WATER HAD POOLED INSIDE AND AROUND DOCKER. THE CUSTOMER WAS ADVISED TO HAVE BUILDING MAINTENANCE REPAIR THE CORRODED FAUCET. THE CORD WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DOCKER'S POWER CORD WAS BURNED AND VISIBLY ARCING. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEPTUNE 2 DOCKING STATION (120V) | FYD | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |