FDA Adverse Event Malfunction Summary report: N

NEPTUNE 2 DOCKING STATION (120V)

MDR report key: 2101945 · Received May 18, 2011

Report

Report Number
1811755-2011-01809
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 18, 2011
Report Date
April 18, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SVC REP EVALUATED THE DOCKER AND FOUND THAT THE FAUCET LOCATED ABOVE THE DOCKER WAS CORRODED AND LEAKING WATER DIRECTLY ONTO THE DOCKER. WATER HAD POOLED INSIDE AND AROUND DOCKER. THE CUSTOMER WAS ADVISED TO HAVE BUILDING MAINTENANCE REPAIR THE CORRODED FAUCET. THE CORD WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DOCKER'S POWER CORD WAS BURNED AND VISIBLY ARCING. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES.