FDA Adverse Event
Malfunction
Summary report: N
ADJ PIN COLLET 2.0-3.2MM
MDR report key: 2101923
·
Received May 18, 2011
Report
- Report Number
- 1811755-2011-01803
- Event Type
- Malfunction
- Date Received
- May 18, 2011
- Date of Event
- January 22, 2011
- Report Date
- January 26, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PIN WAS STICKING IN THE DEVICE DURING ROUTINE MAINTENANCE. THERE WAS NO PT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ PIN COLLET 2.0-3.2MM | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |