FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2101790 · Received May 10, 2011

Report

Report Number
1720753-2011-07134
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 21, 2011
Report Date
May 10, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE SVC REP REPLACED THE LEFT MONITOR AND CLEANED AND REGREASED THE HIGH VOLTAGE CANDLESTICKS. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LEFT MONITOR WOULD LOSE THE LIVE IMAGE DURING FLUOROSCOPY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1