FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2101787 · Received May 10, 2011

Report

Report Number
1720753-2011-07137
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 25, 2011
Report Date
May 10, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP REPLACED THE LEFT MONITOR. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD LOSE THE LIVE IMAGE ON THE LEFT MONITOR DURING FLUOROSCOPIC X-RAY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1