FDA Adverse Event Malfunction Summary report: N

TERUMO CDI 500 BLOOD PARAMETER MONITOR

MDR report key: 2101769 · Received May 10, 2011

Report

Report Number
1828100-2011-01259
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 15, 2011
Report Date
May 10, 2011
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DRY
PMA / PMN Number
K972962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION IN PROGRESS, BUT NOT YET CONCLUDED.

Description of Event or Problem · 1

DURING ROUTINE TESTING OF THE DEVICE AT THE SERVICE CENTER, THE QUALITY ENGINEER REPORTED THAT A FA02 VENOUS MODULE INITIALIZATION FAILURE OCCURRED AT START UP. VISUAL INSPECTION FOUND A BROKEN THERMISTOR. SINCE THE EVENT OCCURRED DURING ROUTINE TESTING, THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO CDI 500 BLOOD PARAMETER MONITOR ON LINE BLOOD GAS MONITOR DRY TERUMO CARDIOVASCULAR SYSTEMS CORP. 500AVHCT