FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2101659 · Received May 11, 2011

Report

Report Number
9617766-2011-01032
Event Type
Malfunction
Date Received
May 11, 2011
Date of Event
April 25, 2011
Report Date
May 11, 2011
Manufacturer
GE MED SYSTEMS (INDIA) PRIVATE LTD
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE CELERON PROCESSOR AND THE HARD DRIVE, AND RELOADED THE SOFTWARE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT DISPLAY AN IMAGE ON EITHER MONITOR. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE MED SYSTEMS (INDIA) PRIVATE LTD 8800

Patients

Seq Age Sex Outcome Treatment
1