FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2101635 · Received May 11, 2011

Report

Report Number
1720753-2011-07221
Event Type
Malfunction
Date Received
May 11, 2011
Date of Event
April 27, 2011
Report Date
May 11, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION AND WAS UNABLE TO DUPLICATE THE REPORTED PROBLEM. THE CABLE AND CARDS RELATED TO THE GENERATOR INTERFACE BOARD AND THE X-RAY GENERATOR WERE RESEATED. THE TUBE WAS REPAIRED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SCREEN OF THE SYSTEM FLICKERED AND WOULD NOT DISPLAY IMAGES INTERMITTENTLY AND THAT THE KILOVOLTAGE WENT TO THE MAXIMUM. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1