FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2101574 · Received May 25, 2011

Report

Report Number
2531779-2011-03659
Event Type
Malfunction
Date Received
May 25, 2011
Date of Event
March 7, 2011
Report Date
April 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). CORRECTION NUMBER: 2531779-03/24/2010-003-R. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED, AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY. THIS REPORT IS BEING MADE BASED ON EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 10