FDA Adverse Event Malfunction Summary report: N

AUTOSOFT 90

MDR report key: 21014862 · Received December 25, 2024

Report

Report Number
3003442380-2024-35198
Event Type
Malfunction
Date Received
December 25, 2024
Date of Event
October 20, 2024
Report Date
November 26, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4), DEVICE 7 OF 12.

Description of Event or Problem · 0

REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN SPAIN. IT WAS REPORTED THAT THE PATIENT FACED TWELVE INFUSION SETS CANNULA KINKED EVENTS ON 09-SEP-2024, 15-SEP-2024, 21-SEP-2024, 30-SEP-2024, 12-OCT-2024, 17-OCT-2024, 20-OCT-2024, 25-OCT-2024, 25-OCT-2024, 31-OCT-2024, 03-NOV-2024, AND 12-NOV-2024. THE EVENT OCCURRED WITHIN THREE HOURS OF INSERTION. THE INSERTION SITE WAS ABDOMEN. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109759 AUTOSOFT 90 UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1002817 6003650

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown