MINICAP TRANSFER SET
Report
- Report Number
- 1423500-2011-06432
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- May 1, 2011
- Report Date
- May 1, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THE PRODUCT CODE IS UNKNOWN. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) THAT OCCURRED DURING DRAIN 12 OF 14 WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE CAUSE WAS UNDETERMINED. THE BATCH REVIEW WAS NOT PERFORMED BECAUSE THE LOT NUMBER IS UNKNOWN. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
THE CAREGIVER (CG) CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOME CHOICE (HC) MACHINE DURING DRAIN 12 OF 14. THE CG STATED THE HOME PATIENT'S (HP) CATHETER BECAME DISCONNECTED AND THE PATIENT'S MOTHER TOOK HIM TO THE HOSPITAL. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THAT THE SE 2240 ALARM INDICATES AIR ENTERED SET UP AND FURTHER ASSISTED THE CG TO CYCLE POWER TO CLEAR THE ALARM. ON (B)(6) 2011, PRODUCT SURVEILLANCE CONTACTED THE HP'S NURSE WHO STATED THE TRANSFER SET SEPARATED FROM THE CATHETER ADAPTER. THE NURSE CONFIRMED THE TRANSFER SET WAS CHANGED OUT AND EVERYTHING WAS FINE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MINICAP TRANSFER SET | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 MO |