HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-06436
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- April 30, 2011
- Report Date
- April 30, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR (SE) 2240 (AIR IN SET) WAS NOT CONFIRMED; HOWEVER, PER THE COMPLAINT INFORMATION THE ROOT CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THERE WAS AN OPEN CLAMP ON UNUSED SUPPLY LINE. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DWELL. THERE WAS AN UNUSED LINE STILL IN THE ORGANIZER AND THE CLAMP WAS OPEN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED TO THE HOME PATIENT (HP) THAT SHE SHOULD ALWAYS CHECK DURING SETUP TO MAKE SURE THAT UNUSED LINES ARE CLAMPED SHUT. THE HP UNDERSTOOD. THE HP TO SETUP WITH NEW SUPPLIES. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE HP. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. DURING A FOLLOW UP WITH THE HP REGARDING THE USE ERROR AND THE ALARM, IT WAS REVEALED THAT THE ISSUE WAS RESOLVED, AND ADDED THAT THE CLAMP WAS ACCIDENTALLY LEFT OPEN. THE HP VERIFIED THAT SHE IS AWARE OF THE PROPER PROCEDURES. PER HP, SHE DID NOT HAVE ANY INJURY OR HARM AS A RESULT OF THIS INCIDENT. THE HP STATED THAT SHE IS DOING FINE AND CONTINUING THERAPY WITHOUT ISSUES. NO FURTHER INFORMATION WAS PROVIDED.