C-MAC VIDEO LARYNGOSCOPE MAC #3
Report
- Report Number
- 9610617-2024-00540
- Event Type
- Malfunction
- Date Received
- December 24, 2024
- Date of Event
- September 1, 2024
- Report Date
- February 3, 2025
- Manufacturer
- KARL STORZ SE & CO. KG
- Product Code
- CCW
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE AFFECTED DEVICE HAS BEEN REQUESTED FOR INVESTIGATION BY THE MANUFACTURER. THE DEVICE WAS NOT YET RETURNED FOR INVESTIGATION. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).
RESULT OF INVESTIGATION: SINCE THE PRODUCT WAS NOT SENT TO KST FOR INSPECTION AND THERE IS NO BATCH NUMBER, NO DETAILED INSPECTION COULD BE CARRIED OUT. BASED ON THE ERROR REPORTED BY THE CUSTOMER (LOOSE CONTACT) AND ON THE BASIS OF COMPARABLE COMPLAINTS, OUR OWN EXPERIENCE HAS LED US TO THE MOST LIKELY CONCLUSION THAT THE ADHESIVE AT THE TIP OF THE C-MAC BLADE MAY HAVE WORN OUT DUE TO THE REMANUFACTURING PROCESS, CAUSING LIQUID TO ENTER THE BLADE, WHICH AFFECTS THE ELECTRONICS AND CAN CAUSE THE LED TO FAIL/DIMINISH IN INTENSITY. FURTHERMORE, THE IMAGE SENSOR CAN BE AFFECTED BY THE INGRESS OF LIQUID AND CAN SHOW IMAGE INTERFERENCE/FAILURE, AND A LOOSE CONTACT CAN ALSO BE THE RESULT. FURTHERMORE, THE CAUSE COULD ALSO BE DROP DAMAGE TO A LOOSE CONTACT OR IT COULD BE A FAULT OF ANOTHER COMPONENT SUCH AS THE CABLE OR SCREEN THAT CAN BE CONNECTED. THE IFU ALSO DESCRIBES THAT A FUNCTIONAL AND VISUAL TEST MUST BE CARRIED OUT BEFORE USE. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).
IT WAS REPORTED THAT DURING THE PROCEDURE THE LIGHT AND THE IMAGE OF THE VIDEO LARYNGOSCOPE FAILED. NO HARM TO PATIENT, USER OR THIRD REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2190926 | C-MAC VIDEO LARYNGOSCOPE MAC #3 | C-MAC VIDEO LARYNGOSCOPE MAC #3 | CCW | KARL STORZ SE & CO. KG | 8404AXC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |