FDA Adverse Event
Malfunction
Summary report: N
QUANTUM MAVERICK BALLOON CATHETER
MDR report key: 2101223
·
Received May 24, 2011
Report
- Report Number
- 2134265-2011-01806
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Report Date
- April 26, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- LOX
- PMA / PMN Number
- P860019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTIONAL TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE QUANTUM MAVERICK MONORAIL 8MM X 3.25MM BALLOON WAS USED FOR PRE-DILATATION. THE BALLOON RUPTURED DURING THE INITIAL INFLATION BELOW THE RATED BALLOON PRESSURE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS' STATUS WAS LISTED AS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | QUANTUM MAVERICK BALLOON CATHETER | CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS | LOX | BOSTON SCIENTIFIC - MAPLE GROVE | H7493808008320 | 0013320988 |