FDA Adverse Event Malfunction Summary report: N

QUANTUM MAVERICK BALLOON CATHETER

MDR report key: 2101223 · Received May 24, 2011

Report

Report Number
2134265-2011-01806
Event Type
Malfunction
Date Received
May 24, 2011
Report Date
April 26, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTIONAL TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE QUANTUM MAVERICK MONORAIL 8MM X 3.25MM BALLOON WAS USED FOR PRE-DILATATION. THE BALLOON RUPTURED DURING THE INITIAL INFLATION BELOW THE RATED BALLOON PRESSURE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS' STATUS WAS LISTED AS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTUM MAVERICK BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493808008320 0013320988

Patients

Seq Age Sex Outcome Treatment
1