FDA Adverse Event Malfunction Summary report: N

UNICEL® DXL 600 ACCESS® IMMUNOASSAY SYSTEM

MDR report key: 2101138 · Received May 24, 2011

Report

Report Number
2122870-2011-01556
Event Type
Malfunction
Date Received
May 24, 2011
Date of Event
April 26, 2011
Report Date
April 26, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K023764
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND THAT THE TUBING IN THE TRANSFER PUMP HAD A CRACK IN IT. THE FSE REPLACED THE WASTE PERI-PUMP TUBING AND VERIFIED THE SYSTEM FOR PROPER OPERATION. HARDWARE IS THE ROOT CAUSE OF THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT A LEAK COMING FROM THE UNICEL DXL 600 ACCESS IMMUNOASSAY SYSTEM FROM FLUIDS TRAY OVER THE LIQUID WASTE COMPARTMENT THE CUSTOMER CONFIRMED THAT PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) WAS WORN WHEN CLEANING UP THE LEAK. NO INJURY HAS BEEN REPORTED IN CONNECTION WITH THIS EVENT.