FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXL 600 ACCESS® IMMUNOASSAY SYSTEM
MDR report key: 2101138
·
Received May 24, 2011
Report
- Report Number
- 2122870-2011-01556
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- April 26, 2011
- Report Date
- April 26, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND THAT THE TUBING IN THE TRANSFER PUMP HAD A CRACK IN IT. THE FSE REPLACED THE WASTE PERI-PUMP TUBING AND VERIFIED THE SYSTEM FOR PROPER OPERATION. HARDWARE IS THE ROOT CAUSE OF THIS EVENT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT A LEAK COMING FROM THE UNICEL DXL 600 ACCESS IMMUNOASSAY SYSTEM FROM FLUIDS TRAY OVER THE LIQUID WASTE COMPARTMENT THE CUSTOMER CONFIRMED THAT PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) WAS WORN WHEN CLEANING UP THE LEAK. NO INJURY HAS BEEN REPORTED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXL 600 ACCESS® IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |