CONTOUR CURVED CUTTER STAPLER
Report
- Report Number
- 3005075853-2011-02116
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- May 4, 2011
- Report Date
- May 4, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K040038
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). DAMAGED CARTRIDGE. THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED WITH THE WASHER UNCUT, WITHOUT STAPLES PRESENT; THIS CONDITION IS CONSISTENT WITH THE DEVICE BEING PARTIALLY ACTIVATED. AS A RESULT OF THE PARTIAL FIRING, THE LOCKOUT WAS ENGAGED WHEN THE DEVICE WAS REOPENED. DO NOT FIRE THE INSTRUMENT UNLESS THE CLOSURE TRIGGER IS PROPERLY LATCHED AGAINST THE HANDLE. THE FIRING TRIGGER MUST BE PULLED BACK COMPLETELY AGAINST THE CLOSURE TRIGGER TO PROPERLY FIRE THE INSTRUMENT. IN ADDITION, IF THE FIRING SEQUENCE IS NOT COMPLETE, YOU COULD DEPLOY THE STAPLES WITHOUT CUTTING THE WASHER AND FORMING THE STAPLES. PLEASE REFERENCE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. IN ADDITION, THE KNIFE AND FOUR DRIVERS WERE FOUND DAMAGED. ONE POSSIBLE CAUSE FOR THE DAMAGED KNIFE MAY BE IF THE DEVICE IS FIRED OVER AN ALREADY EXISTING STAPLE LINE, HARD OBJECT OR THICKER TISSUE THAN INDICATED. REPEATEDLY FIRING ACROSS EXISTING STAPLE LINES CAN ALSO REDUCE THE ABILITY TO CUT CLEANLY. TO MITIGATE THE POTENTIAL FOR STAPLES GETTING INTO THE CARTRIDGE, PROPER CARE SHOULD BE TAKEN WHEN PLACING THE DEVICE ON THE TISSUE TO BE STAPLED THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4). IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. THE BATCH RECORDS WERE REVIEWED AND THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THE BATCHES WERE RELEASED FOR DISTRIBUTION.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, WHEN THE SURGEON CLOSED THE TRIGGER, THERE WAS NO SOUND FEEDBACK. DURING THE FIRING, THE KNIFE COULD ONLY CUT A SMALL PART OF TISSUE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR CURVED CUTTER STAPLER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4UA0R |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE |