ENDOPATH** ETS45
Report
- Report Number
- 3005075853-2011-02107
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Report Date
- May 3, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). SPRING CARTRIDGE LOCKOUT TAB. THE ANALYSIS SHOWED THAT THE ETS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED PARTIALLY FIRED AND WITH THE RELOAD LOCKOUT SPRING DAMAGED. THE DAMAGE TO THE RELOAD LOCKOUT SPRING IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTE TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMA PROCEDURE, THE DEVICE DID NOT STAPLE BUT CUT, NO FURTHER INFORMATION IS AVAILABLE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** ETS45 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | G4U867 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |