FDA Adverse Event Malfunction Summary report: N

ENDOPATH** ETS45

MDR report key: 2100893 · Received May 24, 2011

Report

Report Number
3005075853-2011-02107
Event Type
Malfunction
Date Received
May 24, 2011
Report Date
May 3, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). SPRING CARTRIDGE LOCKOUT TAB. THE ANALYSIS SHOWED THAT THE ETS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED PARTIALLY FIRED AND WITH THE RELOAD LOCKOUT SPRING DAMAGED. THE DAMAGE TO THE RELOAD LOCKOUT SPRING IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTE TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMA PROCEDURE, THE DEVICE DID NOT STAPLE BUT CUT, NO FURTHER INFORMATION IS AVAILABLE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** ETS45 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK G4U867