PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2100888 · Received May 24, 2011

Report

Report Number
2024168-2011-03642
Event Type
Injury
Date Received
May 24, 2011
Date of Event
April 26, 2011
Report Date
May 2, 2011
Manufacturer
AV-REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE ANALYSIS OF THE RETURNED COMPONENTS OF THE DEVICE, HANDLE TO FOOT FUNCTION, GUIDE TUBE, NEEDLE GUIDE, BRIDGE, SUTURE BEARING, EXIT RAMP AND SHEATH WERE NORMAL, THERE WAS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY THAT WOULD CONTRIBUTE TO THE REPORTED CUFF MISS. THE POSTERIOR CUFF WAS STILL LOADED ON THE FOOT. BOTH CUFFS WERE ATTACHED TO THE LINK AND THE ANTERIOR CUFF TABS WERE DAMAGED. BASED ON THE EVIDENCE FOUND ON THE RETURNED DEVICE, THE POSTERIOR CUFF WAS MISSED AND THIS CONFIRMED THE REPORTED CUFF MISS. THE ANTERIOR CUFF DETACHED FROM THE NEEDLE TIP WHICH WAS A DIRECT RESULT OF THE POSTERIOR CUFF MISS. BECAUSE THE NEEDLE DID NOT ENGAGE THE POSTERIOR CUFF, THE CUFF WAS NOT EJECTED FROM THE FOOT. WHEN THE PLUNGER WAS REMOVED FROM THE DEVICE, THE LINK WAS HELD ON ONE END BY THE POSTERIOR CUFF IN THE FOOT POCKET WHILE BEING PULLED ON THE OTHER END BY THE WITHDRAWAL OF THE PLUNGER. THIS RESULTED IN THE ANTERIOR CUFF DETACHING FROM THE NEEDLE. THERE WAS A BEND ON THE POSTERIOR NEEDLE NEAR THE FOLLOWER. IT WAS NOT POSSIBLE TO INSERT THE PLUNGER BECAUSE OF TOO MUCH DRIED BLOOD ON THE NEEDLE LUMEN. THE MOST PROBABLE CAUSE FOR THE POSTERIOR CUFF MISS IS NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO INTERACTION WITH HUMAN TISSUE OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE DURING NEEDLE DEPLOYMENT AS EVIDENCED BY THE NEEDLE STRIKE MARK ON THE POSTERIOR FOOT AND THE BENT ON THE POSTERIOR NEEDLE. A REVIEW OF THE FINISHED DEVICE LOT HISTORY RECORDS DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THERE DOES NOT APPEAR TO BE ANY INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY. TO ASSURE THAT ALL PRODUCTS PERFORM ACCORDING TO SPECIFICATIONS THEY ARE SUBJECT TO INSPECTION DURING MANUFACTURING. IN ADDITION, A QUALITY CONTROL AUDIT INSPECTION IS PERFORMED TO VERIFY PRODUCT QUALITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-REDWOOD CITY 030386H

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention