FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2100763 · Received May 12, 2011

Report

Report Number
1720753-2011-07280
Event Type
Malfunction
Date Received
May 12, 2011
Date of Event
April 29, 2011
Report Date
May 12, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE TRACKING WAS ADJUSTED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS EVENT MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM'S MILLIAMPS WERE OUT OF RANGE ON THE WEEKLY TESTING AND THAT THE TRACKING NEEDED TO BE ADJUSTED. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1