FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2100761 · Received May 24, 2011

Report

Report Number
2531779-2011-03647
Event Type
Malfunction
Date Received
May 24, 2011
Date of Event
March 2, 2011
Report Date
May 2, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #2 DATE OF SUBMISSION 02/12/2015-CORRECTION TO SUSPECT MEDICAL DEVICE SERIAL #.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 07/21/2011. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE AUDIO BOLUS BUTTON HAS BEEN INTERMITTENTLY UNRESPONSIVE FOR THE PAST TWO MONTHS. SHE STATED THAT SHE IS ABLE TO GIVE A BOLUS USING THE OTHER KEYPAD BUTTONS; AND THE PUMP IS NOT EXPOSED TO CLEANING PRODUCTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 43 YR