FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 2100726 · Received May 12, 2011

Report

Report Number
1720753-2011-07241
Event Type
Malfunction
Date Received
May 12, 2011
Date of Event
April 27, 2011
Report Date
May 12, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO CONCLUSION CAN BE DRAWN AS REPAIR INFORMATION IS UNAVAILABLE AND NO ADDITIONAL SERVICE INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM WOULD NOT BOOT UP INTERMITTENTLY AND WOULD NOT DISPLAY AN IMAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1