FDA Adverse Event
Malfunction
Summary report: N
8800
MDR report key: 2100451
·
Received May 13, 2011
Report
- Report Number
- 9617766-2011-01051
- Event Type
- Malfunction
- Date Received
- May 13, 2011
- Date of Event
- May 2, 2011
- Report Date
- May 13, 2011
- Manufacturer
- GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE MONOBLOCK WAS REPLACED, THE FILAMENT WAS CALIBRATED, AND THE COLLIMATOR WAS ALIGNED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A FILAMENT REGULATOR FAILURE ERROR MESSAGE DURING FLUOROSCOPY, AND WOULD NOT DISPLAY AN IMAGE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 8800 | FLUOROSCOPIC X-RAY | JAA | GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD | 8800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |