FDA Adverse Event
Malfunction
Summary report: N
7700
MDR report key: 2100446
·
Received May 13, 2011
Report
- Report Number
- 9680959-2011-01247
- Event Type
- Malfunction
- Date Received
- May 13, 2011
- Date of Event
- May 3, 2011
- Report Date
- May 13, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE BV CAMERA'S FRAME GRABBER IN THE IMAGE PROCESSING COMPUTER WAS REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A "BV CAMERA" ERROR MESSAGE AND HAD AN ISSUE WITH FRAME SYNCHRONIZING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 7700 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |