DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2011-00099
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- May 11, 2011
- Report Date
- May 12, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- JGS
- PMA / PMN Number
- K970330
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED SODIUM RESULTS WAS SAMPLE INTEGRITY. THE QUIKLYTE(R) INTEGRATED MULTISENSOR INSTRUCTIONS FOR USE STATES: "FOLLOW THE INSTRUCTIONS WITH YOUR SPECIMEN COLLECTION DEVICE FOR COLLECTION, TUBE HANDLING, SPIN TIMES, AND G-FORCE ESPECIALLY WHEN USING PLASTIC BLOOD COLLECTION TUBES. FAILURE TO FOLLOW THE BLOOD COLLECTION TUBE INSTRUCTIONS MAY RESULT IN INCREASED MAINTENANCE OR TROUBLESHOOTING OF THE IMT SYSTEM." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED SODIUM (NA) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS WERE REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED ON AN ALTERNATE INSTRUMENT AND HIGHER RESULTS WITHIN THE NORMAL RANGE WERE OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED SODIUM RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | QUIKLYTE® INTEGRATED MULTISENSOR | JGS | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | 1BD853 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |