FDA Adverse Event
Malfunction
Summary report: N
DIMENSION® CLINICAL CHEMISTRY SYSTEM
MDR report key: 2100271
·
Received May 24, 2011
Report
- Report Number
- 2517506-2011-00091
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- May 11, 2011
- Report Date
- May 11, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- MMI
- PMA / PMN Number
- K010313
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT IS WATER QUALITY. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
Description of Event or Problem · 1
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. A REPEAT RUN WAS PERFORMED ON THE SAME SAMPLE AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CARDIAC TROPONIN I FLEX® REAGENT CARTRIDGE | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | CC2026 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |