FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2100263 · Received May 16, 2011

Report

Report Number
1644487-2011-01094
Event Type
Malfunction
Date Received
May 16, 2011
Date of Event
January 1, 2010
Report Date
April 18, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

DURING MANUFACTURER PRODUCT ANALYSIS OF AN EXPLANTED VNS GENERATOR, IT WAS NOTED THE GENERATOR WAS SET TO DEFAULT SYSTEMS DIAGNOSTICS SETTINGS OF 1MA/20HZ/500 PULSEWIDTH/30 SEC ON/60 MIN OFF. THIS IS LIKELY DUE TO FAULTED SYSTEMS DIAGNOSTIC TEST EITHER AT THE TIME OF GENERATOR EXPLANT ON (B)(6) 2010, OR PRIOR TO THIS DATE. ATTEMPTS FOR FURTHER INFORMATION AND VNS PROGRAMMING HISTORY ARE IN PROGRESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250

Patients

Seq Age Sex Outcome Treatment
1 26 YR