FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2100257 · Received May 16, 2011

Report

Report Number
9617766-2011-01056
Event Type
Malfunction
Date Received
May 16, 2011
Date of Event
May 6, 2011
Report Date
May 16, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CMOS SETTINGS WERE RECONFIGURED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP AND THE MONITOR'S GREEN LIGHT CAME ON BUT THE MONITOR WOULD NOT POWER UP. THE SYSTEM DISPLAYED CMOS ERROR MESSAGES. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1