FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2100252
·
Received May 16, 2011
Report
- Report Number
- 1720753-2011-07364
- Event Type
- Malfunction
- Date Received
- May 16, 2011
- Date of Event
- May 4, 2011
- Report Date
- May 16, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE COLLIMATOR CABLES WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A "COLLIMATOR IRIS IS TOO LARGE" ERROR MESSAGE. NO PT INJURY WAS REPORTED.