FDA Adverse Event Malfunction Summary report: N

ARCHITECT HAVAB IGM REAGENT

MDR report key: 2100169 · Received May 24, 2011

Report

Report Number
2623532-2011-00016
Event Type
Malfunction
Date Received
May 24, 2011
Report Date
April 14, 2011
Manufacturer
ABBOTT DIAGNOSTICS INT'L, LTD
Product Code
LOL
PMA / PMN Number
K063329
Removal / Correction Number
2623532-1/4/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4), CONCOMITANT MEDICAL DEVICE: ARCHITECT I2000SR ANALYZER, LIST # 3M74-95, SERIAL # (B)(4). EVALUATION RESULTS: CROSS CONTAMINATION WAS IDENTIFIED AS A POTENTIAL CAUSE. CONCLUSION: INSUFFICIENT ACTIVE ANTIGEN IS BEING COATED ON THE MICROPARTICLE WITH THE CURRENT MICROPARTICLE MANUFACTURING PROCESS. AN ADVERSE TREND IN US CUSTOMER COMPLAINTS FOR ARCHITECT (B)(6) ASSAY (LIST NUMBER 6L21) REGARDING HIGHER THAN USUAL (B)(6) RESULTS RATE WAS DETECTED ON (B)(6) 2009. THE INVESTIGATION REVEALED THE MOST PROBABLE CAUSE FOR THE INCREASE RATE OF (B)(6) RESULTS IS THE (B)(6) WHICH HAS A REDUCED POTENCY THAT AFFECTS THE ARCHITECT (B)(6) PERFORMANCE. INSUFFICIENT ACTIVE ANTIGEN IS BEING COATED ON THE MICROPARTICLE WITHIN THE CURRENT MICROPARTICLE MANUFACTURING PROCESS RESULTING IN A PERFORMANCE THAT IS CHARACTERIZED BY LOWER CALIBRATOR 1 RELATIVE LIGHT UNITS (RLU) VALUES. AS A RESULT, THE SIGNAL TO NOISE RATIO IS SHIFTED SPECIALLY IN THE NEGATIVE SAMPLES LEAVING THE ASSAY PRONE FOR (B)(6) RESULTS AND INCREASES ARCH (B)(6) ASSAY SUSCEPTIBILITY TO REAGENT CROSS CONTAMINATION AND TEST SYSTEM VARIABILITY. AS A PREVENTIVE MEASURE TO PROTECT THE INTEGRITY OF TEST RESULTS, CUSTOMERS WERE INSTRUCTED BY A PRODUCT INFORMATION LETTER TO FOLLOW AN ALTERNATIVE PROCESSING METHOD BY SEGREGATING ALL (B)(6) SAMPLES.

Description of Event or Problem · 1

THE CUSTOMER NOTICED AN INCREASE IN INITIAL REACTIVE RATES FOR ARCHITECT (B)(6) PATIENT SAMPLES. THE CUSTOMER RECALIBRATED THE ASSAY ON THE ANALYZER WHICH DID NOT RESOLVE THE ISSUE. THE CUSTOMER WAS ADVISED TO PERFORM VARIOUS TROUBLESHOOTING PROCEDURES WHICH RESOLVED THE ISSUE. THE MOST LIKELY CAUSE OF THE INITIAL REACTIVE RATES OBSERVED BY THE CUSTOMER WAS DUE TO THE POOR CONDITION OF THE ANALYZER PROBE OR THE REAGENT. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT HAVAB IGM REAGENT FOR THE DETECTION OF IGM ANTIBODIES TO HEPATITIS A VIRUS LOL ABBOTT DIAGNOSTICS INT'L, LTD

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT I2000SR ANALYZER, LIST # 3M74-95