FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK

MDR report key: 20999830 · Received December 23, 2024

Report

Report Number
1213809-2024-00988
Event Type
Malfunction
Date Received
December 23, 2024
Date of Event
November 27, 2024
Report Date
January 17, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B.3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT. NO ROOT CAUSE CAN BE DETERMINED AS NO SAMPLES WERE RECEIVED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. WE WOULD BE VERY INTERESTED IN EXAMINING PRODUCT THAT DOES NOT MEET YOUR EXPECTATIONS AND OUR QUALITY STANDARDS. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. EXAMINATION OF THE PRODUCT INVOLVED MAY PROVIDE CLARIFICATION AS TO THE CAUSE FOR THE REPORTED FAILURE. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

MATERIALS: 309628 BATCH#: 1110009 IT WAS REPORTED THAT THE BD LUER-LOK HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. 1 ML SYRINGE (P/N BD-309628, LOT 1110009) WE HAVE IDENTIFIED THE FOLLOWING PARTICULATES IN OUR FINAL DRUG PRODUCT: - CELLULOSE/LIGNIN - CELLULOSE FIBER - POLYPROPYLENE - POLYETHYLENE - PVDF-HFP - TEXTILE COTTON FIBER - TEXTILE POLYESTER FIBER - EVA - INDIGO CELLULOSE OR COTTON FIBER - FLUOROCARBON-BASED MATERIAL.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
80799 BD LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 1110009 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown