FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM

MDR report key: 20999756 · Received December 23, 2024

Report

Report Number
0002023141-2024-04143
Event Type
Injury
Date Received
December 23, 2024
Date of Event
September 29, 2024
Report Date
April 7, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K061410
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER: (B)(4). G4: PMA/510(K) NUMBER: K061410, K011028 AND K013227.

Additional Manufacturer Narrative · 0

ZIMVIE RECEIVED ONE (1) TSVWB8, (IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE, APPARENT BONE / TISSUE ATTACHED TO EXTERNAL THREADS. A SLIGHT FRACTURE WAS IDENTIFIED AT THE IMPLANT'S COLLAR. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENTS COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1233652. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENTS, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1233652 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿FRACTURE IMPLANT¿ & ¿PERI-IMPLANTITIS¿ THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENTS. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW AS PER RMF RM-00541-HAZ REV. 4, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE PARAFUNCTIONAL HABITS/PATIENT FACTORS. REFER TO ATTACHED SUMMARY INVESTIGATION FOR ¿PERI-IMPLANTITIS¿ THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT (FRACTURE IMPLANT) WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. PERI-IMPLANTITIS IS A MEDICAL CONDITION AND IS NON-VERIFIABLE WITH THE INFORMATION PROVIDED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. PERI-IMPLANTITIS IS AN INFECTIOUS DISEASE THAT CAUSES INFLAMMATION OF THE GUM AND THE BONE STRUCTURE AROUND A DENTAL IMPLANT. CHRONIC INFLAMMATION CAUSES BONE LOSS, WHICH CAN LEAD TO IMPLANT FAILURE THUS ITS REMOVAL. INVESTIGATIONS PERFORMED FOR HUNDREDS OF EVENTS WHERE EITHER OF THESE CONDITIONS, INFECTION, OR BONE LOSS, OR BOTH, HAVE BEEN REPORTED, HAVE CONCLUDED THAT THE LIKELY CAUSE(S) FOR THE IMPLANT FAILURE IN RELATION TO THESE CONDITIONS ARE EXTERNAL FACTORS. THOSE INCLUDE MEDICAL CONDITIONS (E.G., DIABETES, BRUXISM, ETC.), PATIENT HABITS (E.G., SMOKING, BAD ORAL HYGIENE ROUTINE) AND USER ERROR (E.G., SURGICAL TECHNIQUE). THERE HAS NOT BEEN ANY INSTANCE WHERE EITHER INFECTION AND/OR BONE LOSS, AND WHEN RIGHT CONDITIONS PREVAIL AND LED TO PERI-IMPLANTITIS WHETHER REPORTED OR NOT ALONG WITH THE OTHER TWO, HAVE RESULTED FROM A MANUFACTURING OR DESIGN DEFECT. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO INFECTION AND RESULT IN BONE LOSS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THE ANALYSIS AND RESULT OF INVESTIGATIONS AND THE ANALYSIS OF PROBABILITY PREVIOUSLY DESCRIBED ARE CONTAINED IN THE SUMMARY INVESTIGATION REPORTS PERFORMED FOR BONE LOSS AND INFECTION, WHICH ARE ATTACHED. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/ OR ANY FURTHER ESCALATIONS. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

PER INDICATED: FRACTURE, LOSS INTEGRATION, PERI-IMPLANTITIS. THE PATIENT WAS EXAMINED AND FOUND TO HAVE A FRACTURE AND LACK OF OCCLUSAL FUNCTION, THE PATIENT UNDERWENT IMPLANT SURGERY ON (B)(6) 2024 IN THE UPPER RIGHT:26 POSITION USING A TEMPORARY ABUTMENT, IMPLANT MODEL NUMBER: TSVWB8. ON (B)(6) 2024, AN EXAMINATION REVEALED A HIGH DEGREE OF PERI-IMPLANT SHADOWING, AND SURGERY WAS PERFORMED TO REMOVE THE IMPLANT. SYMPTOMS AS A RESULT OF THE EVENT: NONE. SURGICAL/MEDICAL INTERVENTION REQUIRED FOR PERMANENT DAMAGE: NO. ADDITIONAL APPOINTMENT REQUIRED: NO. WAS THE PROCEDURE COMPLETED USING ANOTHER IMPLANT OR ANOTHER DEVICE? NO.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
77633 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM DENTAL IMPLANT DZE ZIMMER DENTAL 1233652

Patients

Seq Age Sex Outcome Treatment
1 38 YR Female Required Intervention