IMPL TAPERED SCR-V HA 6MM 5.7MM 11.5MM
Report
- Report Number
- 0002023141-2024-04137
- Event Type
- Injury
- Date Received
- December 23, 2024
- Date of Event
- November 26, 2024
- Report Date
- April 6, 2025
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. G4: PMA/510(K) NUMBER: K061410 AND K013227.
ZIMVIE RECEIVED ONE (1) TSV6H11, (IMPL TAPERED SCR-V HA 6MM 5.7MM 11.5MM) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE, APPARENT BONE / TISSUE ATTACHED TO EXTERNAL THREADS. IMPLANT FRACTURE IDENTIFIED AT THE COLLAR. DEBRIS WERE NOTED INSIDE THE IMPLANT. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1260108. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1260108 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿FRACTURE IMPLANT¿. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW AS PER RMF RM-00541-HAZ REV. 4, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE PARAFUNCTIONAL HABITS/PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #30 WAS REMOVED AS IMPLANT WAS FRACTURED. PATIENT HAD HISTORY OF CROWN GETTING LOOSE ON 2 SEPARATE OCCASIONS. FOUND THAT IMPLANT WAS FRACTURED AND REMOVED IMPLANT. PER INDICATED: SURGICAL/MEDICAL INTERVENTION REQUIRED FOR PERMANENT DAMAGE: NO. WAS THE PROCEDURE COMPLETED USING ANOTHER IMPLANT OR ANOTHER DEVICE: YES. ADDITIONAL APPOINTMENT REQUIRED: NO. SYMPTOMS AS A RESULT OF THE EVENT: N/A.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2625922 | IMPL TAPERED SCR-V HA 6MM 5.7MM 11.5MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1260108 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Male | Required Intervention |