FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK

MDR report key: 20996856 · Received December 21, 2024

Report

Report Number
1213809-2024-00980
Event Type
Malfunction
Date Received
December 21, 2024
Date of Event
November 5, 2024
Report Date
February 18, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096573
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. DURING THE SAMPLE EVALUATION, MULTIPLE UNKNOWN BROWN SPOTS WERE OBSERVED TO BE EMBEDDED IN THE BARREL. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. THE POTENTIAL ROOT CAUSE FOR THE EMBEDDED FOREIGN MATTER DEFECT IS ASSOCIATED WITH THE MOLDING PROCESS. THE LOT NUMBER IS UNKNOWN, THEREFORE DEVICE HISTORY RECORD REVIEW (DHR) OR QUALITY NOTIFICATION REVIEW (QN) COULD NOT BE PERFORMED. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

B.3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

MATERIAL: 309657. BATCH#: UNKNOWN (PROVIDED LOT #4102973). IT WAS REPORTED THAT THE BD LUER-LOK HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. DARK STAINED COLOR ON SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54664 BD LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS UNKNOWN 30382903096573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown