ENDOWRIST
Report
- Report Number
- 2955842-2024-23547
- Event Type
- Malfunction
- Date Received
- December 21, 2024
- Date of Event
- November 26, 2024
- Report Date
- November 26, 2024
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE SMALL GRASPING RETRACTOR INSTRUMENT WAS ANALYZED AND FOUND TO HAVE THE CABLE CRIMP DISLODGED AT THE DISTAL END. THE COMPLAINT REGARDING THE HINGE SPRING BROKE WAS CONFIRMED BY FAILURE ANALYSIS.
INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE SMALL GRASPING RETRACTOR INSTRUMENT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS; HOWEVER, THE EVALUATION IS NOT YET COMPLETE.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CHOLECYSTECTOMY SURGICAL PROCEDURE, THE 8MM SMALL GRASPING RETRACTOR INSTRUMENT'S HINGE SPRING BROKE. NO FRAGMENTS FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED BUT THE PATIENT OUTCOME IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2327664 | ENDOWRIST | SMALL GRASPING RETRACTOR | NAY | INTUITIVE SURGICAL, INC | 470318-10 | K10240829 0235 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |