FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 20996160 · Received December 21, 2024

Report

Report Number
2955842-2024-23533
Event Type
Malfunction
Date Received
December 21, 2024
Date of Event
November 19, 2024
Report Date
December 3, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
PMA / PMN Number
K182371
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN MOLDED INSULATOR. AS A RESULT, THE GRIP TIP WAS DISLODGED. A BROKEN PIECE APPROXIMATELY 3.35MM X 9.55MM IN SIZE WAS MISSING AND NOT RETURNED WITH THE INSTRUMENT. THE COMPLAINT REGARDING THE BROKEN DEVICE WAS CONFIRMED BY FAILURE ANALYSIS BLANK MDR FIELDS: THE MISSING PATIENT INFORMATION IN SECTIONS A AND B WAS EITHER UNKNOWN, UNAVAILABLE, NOT PROVIDED, OR NOT APPLICABLE. THE EXPIRATION DATE FOR SECTION D4 IS NOT APPLICABLE. FIELD D6 IS BLANK BECAUSE THE PRODUCT IS NOT IMPLANTABLE. INFORMATION FOR THE BLANK FIELDS IN SECTION E1 IS NOT AVAILABLE. FIELD E4 IS BLANK BECAUSE IT IS UNKNOWN IF THE INITIAL REPORTER SUBMITTED A REPORT TO THE FDA. FIELDS G5 AND G7 ARE NOT APPLICABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING DA VINCI-ASSISTED SURGICAL PROCEDURE, TIP OF FENESTRATED BIPOLAR FORCEPS WAS FOUND TO BE BROKEN. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE DAMAGE THAT OCCURRED DID NOT OCCUR WHILE ARM WAS IN USE IN THE PATIENT. NO FRAGMENTS WERE REPORTED TO HAVE FALLEN INTO PATIENT AS IT DID NOT HAPPEN IN SURGICAL SITE. THE PATIENT HAS NOT RETURNED TO THE HOSPITAL DUE TO ANY COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2327661 ENDOWRIST SP FENESTRATED BIPOLAR FORCEPS NAY INTUITIVE SURGICAL, INC 430011-57 K10240229 0029

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES