NEEDLE NS 25GA 3/4IN W/O SILICONE
Report
- Report Number
- 1911916-2024-00928
- Event Type
- Malfunction
- Date Received
- December 20, 2024
- Date of Event
- December 9, 2024
- Report Date
- January 16, 2025
- Manufacturer
- BECTON DICKINSON
- Product Code
- FMI
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PR (B)(4) FOLLOW UP. IT WAS REPORTED THE CYSTOTOME IN OUR CUSTOM PACKS HAVE BEEN COMING WITH A SUB-CYSTOTOME, AND THE WHITE PLASTIC LOOKS LIKE IT MELTED ON THE METAL. TO AID IN THE INVESTIGATION, TWO PHOTOS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTOS SHOW A NEEDLE ASSEMBLY WITH AN EPOXY DRIP OVER ON THE NEEDLE HUB. THE NEEDLE IS BENT AS PART OF THE PROCESSING DONE BY THE DISTRIBUTOR. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THE EPOXY ON THE NEEDLE HUB COULD OCCUR IF THERE WAS A JAM AT THE CANNULATOR. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 301670, LOT 3053274. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECTS. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. VERIFICATION OF THE CANNULATOR WAS PERFORMED. THE SETTINGS WERE CORRECT, AND THE FLOW OF PRODUCT WAS GOOD. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.
INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
NO ADDITIONAL INFORMATION RECEIVED.
MATERIAL#: 301670 BATCH NUMBER#: 3053274IT WAS REPORTED BY CUSTOMER THAT THE CYSTOTOME IN OUR CUSTOM PACKS HAVE BEEN COMING WITH SUB CYSTOTOME. TODAY THE WHITE PLASTIC LOOKS LIKE IT MELTED ON METAL. THEY ARE PULLING ONE ON THE SIDE IF THEY SEE THIS. NO HARM
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2045115 | NEEDLE NS 25GA 3/4IN W/O SILICONE | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | BECTON DICKINSON | 3053274 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |