FDA Adverse Event Malfunction Summary report: N

SIGMA SPECTRUM INFUSION PUMP

MDR report key: 20988052 · Received December 20, 2024

Report

Report Number
20988052
Event Type
Malfunction
Date Received
December 20, 2024
Date of Event
October 9, 2024
Report Date
November 18, 2024
Manufacturer
BAXTER HEALTHCARE CORPORATION.
Product Code
PHC
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

PATIENT HAD PRECEDEX INFUSION INITIATED IN POST ANESTHESIA CARE UNIT (PACU). PUMP WAS SET TO DELIVER 3.4 MILLILITERS PER HOUR AND WAS RUNNING FOR 45 MINUTES. NURSE REPORTED THAT MEDICATION BAG WAS DRY AND THEN QUESTIONED HOW A 100 MILLILITER BAG WAS DRY IN 45 MINUTES. PUMP SAID TOTAL GIVEN WAS 2 MILLILITERS. OPERATING ROOM (OR) PHARMACIST CAME TO BEDSIDE AND ENSURED THAT MEDICATION BAG HAD 100ML WHEN IT WAS DISPENSED. ANESTHESIA AND DOCTOR¿S TEAM NOTIFIED. PATIENT VITALLY STABLE BUT SEDATED. INTRAVENOUS (IV) PUMP TAKEN OUT OF SERVICE AND MARKED FOR REPAIR. PUMP WAS TESTED BY OUR HEALTH TECHNOLOGY MANAGMENT (HTM), WHO STATED THEY SENT THE PUMP TO BAXTER FOR REPAIR. THEIR TESTING: RAN 10% OVER ON A COUPLE OF DIFFERENT INFUSIONS. MANUFACTURER RESPONSE FOR INFUSION SAFETY MANAGEMENT SOFTWARE, SIGMA SPECTRUM INFUSION PUMP (PER SITE REPORTER).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2204240 SIGMA SPECTRUM INFUSION PUMP INFUSION SAFETY MANAGEMENT SOFTWARE PHC BAXTER HEALTHCARE CORPORATION. 35700BAX2

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown