FDA Adverse Event Injury Summary report: N

STELLANT DUAL CT INJECTOR

MDR report key: 20986997 · Received December 20, 2024

Report

Report Number
2520313-2024-00023
Event Type
Injury
Date Received
December 20, 2024
Date of Event
November 30, 2024
Report Date
December 20, 2024
Manufacturer
BAYER MEDICAL CARE INC
Product Code
DXT
PMA / PMN Number
K192370
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SYSTEM SERVICE CHECK OF THE MEDRAD® STELLANT FLEX INJECTION SYSTEM, SERIAL NUMBER (B)(6), WAS PERFORMED ON (B)(6) 2024, WHICH CONFIRMED THAT THE INJECTOR WAS OPERATING WITHIN BAYER SPECIFICATIONS. THE DISPOSABLES THAT WERE IN USE DURING THE PROCEDURE WERE DISCARDED BY THE SITE; THEREFORE, THEY ARE NOT AVAILABLE FOR EVALUATION. THE CUSTOMER WAS UNABLE TO PROVIDE THE LOT NUMBER OF THE DISPOSABLES USED DURING THE PROCEDURE; THEREFORE, TESTING OF RETAINED SAMPLES WAS NOT POSSIBLE. THE OFFER OF ADDITIONAL CLINICAL APPLICATIONS TRAINING WAS ACCEPTED BY THE CUSTOMER AND WAS SCHEDULED WITH THE CLINICAL PERFORMANCE CENTER. THE MEDRAD® STELLANT FLEX INJECTION SYSTEM OPERATION MANUAL CAUTIONS THE USER AS FOLLOWS: WARNING: AIR EMBOLISM HAZARD - SERIOUS PATIENT INJURY OR DEATH MAY RESULT. ENSURE PATIENT IS NOT CONNECTED WHILE PURGING AIR FROM SYRINGE, OR ENGAGING OR ADVANCING PLUNGER. EXPEL ALL TRAPPED AIR FROM THE SYRINGE(S), CONNECTORS, TUBING, AND CATHETER BEFORE CONNECTING THE SYSTEM TO THE PATIENT. TO MINIMIZE AIR EMBOLIZATION RISKS, ENSURE THAT ONE OPERATOR IS DESIGNATED THE RESPONSIBILITY OF FILLING THE SYRINGE(S). DO NOT CHANGE OPERATORS DURING THE PROCEDURE. IF AN OPERATOR CHANGE MUST OCCUR, ENSURE THAT THE NEW OPERATOR VERIFIES THAT THE FLUID PATH IS PURGED OF AIR. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT.

Description of Event or Problem · 0

BAYER MEDICAL CARE INC. WAS INFORMED THAT A PATIENT HAD EXPERIENCED AN ALLEGED AIR INJECTION WHILE CONNECTED TO A MEDRAD® STELLANT FLEX INJECTION SYSTEM (SN (B)(6)). THE AMOUNT AND ANATOMICAL LOCATION OF THE AIR WAS UNDISCLOSED. FOLLOWING THE INJECTION, THE PATIENT WAS RETURNED TO THE EMERGENCY DEPARTMENT AND WAS PLACED IN A DECUBITUS POSITION. A RADIOLOGIST NOTIFIED THE EMERGENCY DEPARTMENT STAFF OF THE AIR EMBOLUS, AND THE PATIENT WAS SUBSEQUENTLY TRANSFERRED TO THE INTENSIVE CARE UNIT FOR MONITORING. THE PATIENT IS REPORTED AS DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2062762 STELLANT DUAL CT INJECTOR CT INJECTION SYSTEM DXT BAYER MEDICAL CARE INC 85631829

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other