FDA Adverse Event Malfunction Summary report: N

BD NEXIVA SINGLE PORT

MDR report key: 20984137 · Received December 19, 2024

Report

Report Number
1710034-2024-01523
Event Type
Malfunction
Date Received
December 19, 2024
Date of Event
December 6, 2024
Report Date
January 31, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
PMA / PMN Number
K183399
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: THE COMPLAINT OF A LEAK FROM THE SEPTUM WAS CONFIRMED AND THE CAUSE APPEARED TO BE MANUFACTURING RELATED. FOUR 18G NEXIVA UNITS FROM LOT 4232063 WERE PROVIDED FOR INVESTIGATION. THE CANNISTER AND SEPTUM WERE NOT SEATED CORRECTLY WITHIN THE CATHETER ADAPTER, WHICH LIKELY CREATED A FLUID PATH THAT ALLOWED FLUID TO LEAK. CORRECTIVE ACTIONS HAVE BEEN INITIATED, AND A MANUFACTURING ROOT CAUSE HAS BEEN IDENTIFIED FOR THIS FAILURE MODE. ACTIONS HAVE BEEN MADE TO ADDRESS THE ISSUE. A FUNCTIONAL TEST SHOWED THAT NO FLUID WOULD BYPASS THE CLOSED CLAMP; THEREFORE, THE COMPLAINT OF BACKFLOW COULD NOT BE CONFIRMED. THE APPROPRIATE MANUFACTURING PERSONNEL WERE NOTIFIED OF THIS COMPLAINT. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD NEXIVA SINGLE PORT CLAMP DOES NOT STOP BLOOD BACKFLOW. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INITIAL IV STARTED ON PATIENT MALFUNCTIONED. SECOND IV PLACEMENT WAS SUCCESSFUL, HOWEVER THE CLAMP ON THE IV DOES NOT KEEP BLOOD FROM BACKFLOWING, WHICH PUTS IT AT A HIGH RISK FOR CLOTTING OFF. THIS HAPPENS REGULARLY, HOWEVER ON THIS INDIVIDUAL WE NOW HAVE VERY FEW INSERTION SITES LEFT AS SHE HAS A RESTRICTED EXTREMITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2202438 BD NEXIVA SINGLE PORT PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4232063

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown