BD NEXIVA SINGLE PORT
Report
- Report Number
- 1710034-2024-01523
- Event Type
- Malfunction
- Date Received
- December 19, 2024
- Date of Event
- December 6, 2024
- Report Date
- January 31, 2025
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- PMA / PMN Number
- K183399
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- 003
Narratives
INVESTIGATION RESULTS: THE COMPLAINT OF A LEAK FROM THE SEPTUM WAS CONFIRMED AND THE CAUSE APPEARED TO BE MANUFACTURING RELATED. FOUR 18G NEXIVA UNITS FROM LOT 4232063 WERE PROVIDED FOR INVESTIGATION. THE CANNISTER AND SEPTUM WERE NOT SEATED CORRECTLY WITHIN THE CATHETER ADAPTER, WHICH LIKELY CREATED A FLUID PATH THAT ALLOWED FLUID TO LEAK. CORRECTIVE ACTIONS HAVE BEEN INITIATED, AND A MANUFACTURING ROOT CAUSE HAS BEEN IDENTIFIED FOR THIS FAILURE MODE. ACTIONS HAVE BEEN MADE TO ADDRESS THE ISSUE. A FUNCTIONAL TEST SHOWED THAT NO FLUID WOULD BYPASS THE CLOSED CLAMP; THEREFORE, THE COMPLAINT OF BACKFLOW COULD NOT BE CONFIRMED. THE APPROPRIATE MANUFACTURING PERSONNEL WERE NOTIFIED OF THIS COMPLAINT. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY.
H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
NO ADDITIONAL INFORMATION.
IT WAS REPORTED THAT BD NEXIVA SINGLE PORT CLAMP DOES NOT STOP BLOOD BACKFLOW. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INITIAL IV STARTED ON PATIENT MALFUNCTIONED. SECOND IV PLACEMENT WAS SUCCESSFUL, HOWEVER THE CLAMP ON THE IV DOES NOT KEEP BLOOD FROM BACKFLOWING, WHICH PUTS IT AT A HIGH RISK FOR CLOTTING OFF. THIS HAPPENS REGULARLY, HOWEVER ON THIS INDIVIDUAL WE NOW HAVE VERY FEW INSERTION SITES LEFT AS SHE HAS A RESTRICTED EXTREMITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2202438 | BD NEXIVA SINGLE PORT | PERIPHERAL IV CATHETERS | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 4232063 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |