FDA Adverse Event Summary report: N

TSRH SPINAL SYSTEM

MDR report key: 2097749 · Received May 20, 2011

Report

Report Number
1030489-2011-00611
Date Received
May 20, 2011
Date of Event
April 26, 2011
Report Date
August 5, 2011
Manufacturer
WARSAW ORTHOPEDIC, INC.
Product Code
KWP
PMA / PMN Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS PART IS NOT APPROVED FOR USE IN THE UNITED STATES; HOWEVER A LIKE DEVICE CATALOG # 8372645, 510K # K984522 WAS CLEARED IN THE UNITED STATES. NEITHER DEVICE NOR FILM OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT. DEVICE HISTORY RECORDS FOR THIS LOT WERE REVIEWED. NO DOCUMENTATION WAS FOUND THAT WOULD INDICATE A NON-CONFORMANCE TO SPECIFICATIONS.

Additional Manufacturer Narrative · 1

THE SCREW WAS RETURNED FOR ANALYSIS. MACROSCOPIC EXAMINATION CONFIRMS BONE SCREWS ARE BROKEN APPROX. ~4-8 THREADS FROM THE START OF THE SCREW THREAD. MICROSCOPIC EXAMINATION OF EACH OF THE FRACTURE SURFACES REVEAL A FAIRLY FLAT FRACTURE, WITH PROGRESSIVE STRIATIONS, INDICATIVE OF FATIGUE, FOLLOWED BY ULTIMATE FAILURE OF THE IMPLANT, CONSISTENT WITH FAILURE DUE TO FATIGUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT WAS FUSED FOUR YEARS AFTER A L4-S1 SPINAL PROCEDURE. THE REMOVAL SURGICAL PROCEDURE WAS PERFORMED TO EXPLANT THE POSTERIOR FIXATION SYSTEM. IT WAS FOUND DURING THE REMOVAL PROCEDURE THAT THE PEDICLE SCREWS AT RIGHT L4 (6.5MM) AND BOTH SIDE OF S1 (RIGHT 5.5MM, LEFT 6.5MM) HAD BEEN BROKEN. REPORTEDLY THE RIGHT S1 SCREW WAS CONFIRMED BROKEN THREE YEARS POST OP. TIPS OF THOSE BROKEN SCREWS REMAINED IN THE PATIENT. NO OTHER COMPLICATION WAS REPORTED DURING AND AFTER THE REMOVAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TSRH SPINAL SYSTEM KWP WARSAW ORTHOPEDIC, INC. NA W06J3173

Patients

Seq Age Sex Outcome Treatment
1 00037 YR Other