FDA Adverse Event Malfunction Summary report: N

NEEDLE 18X1-1/2 RB

MDR report key: 20976792 · Received December 19, 2024

Report

Report Number
1911916-2024-00924
Event Type
Malfunction
Date Received
December 19, 2024
Date of Event
December 6, 2024
Report Date
January 14, 2025
Manufacturer
BECTON DICKINSON
Product Code
FMI
UDI-DI
30382903051961
PMA / PMN Number
K021475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP. IT WAS REPORTED THERE WERE TINY HOLES ON NEEDLE SHAFT. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 305196 AND LOT NUMBER 3299400. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. THE ACTUAL DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER WAS ENTERED INTO THE DATE OF EVENT FIELD.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED

Description of Event or Problem · 0

MATERIAL#: 305196, BATCH NUMBER#: 3299400. FIVE TINY HOLES ON NEEDLE SHAFT CAUSING LEAKAGE AND MEDICATION WASTE WHILE DRAWING UP MEDICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2098177 NEEDLE 18X1-1/2 RB NEEDLE, HYPODERMIC, SINGLE LUMEN FMI BECTON DICKINSON 3299400 30382903051961

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown