FDA Adverse Event Injury Summary report: N

IMP,TSV,MCOL MG,3.7MM,11.5MML

MDR report key: 20976303 · Received December 19, 2024

Report

Report Number
0002023141-2024-04100
Event Type
Injury
Date Received
December 19, 2024
Date of Event
July 11, 2024
Report Date
May 12, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K111889
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1233257. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1233257 FOR SIMILAR EVENTS AND TWO OTHER COMPLAINTS WERE IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: BONE LOSS. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). D9: DEVICE AVAILABILITY UNKNOWN / NOT PROVIDED D10. CONCOMITANT MEDICAL PRODUCTS TSVMB11, IMP, TSV, MCOL MG,3.7MM,11.5MML LOT 1233257 AND TSVMB10, IMP, TSV, MCOL MG,3.7MM,10M LOT 1234903. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

DOCTOR REPORTED THAT IMPLANTS AT TOOTH SITES 33, 36, 46 WERE REMOVED AFTER THREE AND A HALF YEARS THE IMPLANTS WERE PLACED AND PROSTHETIC RESTORED DUE TO BONE LOSS. DENTAL PROSTHESIS ON LOCATORS STARTED TO OSCILATE. PATIENT IS ON A SUBSTITUTIONAL THERAPY FOR DIABETES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2099089 IMP,TSV,MCOL MG,3.7MM,11.5MML DENTAL IMPLANT DZE ZIMMER DENTAL 1233257

Patients

Seq Age Sex Outcome Treatment
1 76 YR Female DENTAL IMPLANT SEE H10 NARRATIVE| DENTAL IMPLANT SEE H10 NARRATIVE