FDA Adverse Event Malfunction Summary report: N

ANES CIRCUITS

MDR report key: 2097226 · Received May 11, 2011

Report

Report Number
2242551-2011-00003
Event Type
Malfunction
Date Received
May 11, 2011
Date of Event
April 13, 2011
Report Date
May 11, 2011
Manufacturer
VITAL SIGNS, INC.
Product Code
CAI
PMA / PMN Number
EXEMPT
Removal / Correction Number
2242551-04/29/11-004-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

NO PT INFO HAS BEEN MADE AVAILABLE. THE HME FILTER WAS RETURNED AND EVALUATED BY GE HEALTHCARE ENGINEERS. SAMPLES FROM IN-HOUSE INVENTORY WERE ALSO DRAWN PER A SAMPLING PLAN AND EVALUATED. IT WAS DETERMINED THAT A PLASTIC MEMBRANE OCCLUDED THE MAJORITY OF THE INTERNAL OPENING OF THE FILTER. THE ROOT CAUSE IS BELIEVED TO BE THAT THE POSITION OF THE MOLD AND THE TORQUE APPLIED TO THE MOLD DURING THE PLASTIC HOUSING MFG WERE NOT SET PROPERLY RESULTING IN A REMNANT PLASTIC MEMBRANE. IT WAS DETERMINED THAT THE PART INSPECTION PROCESS WAS NOT ADEQUATE TO DETECT THIS DEFECT. THE AFFECTED PRODUCTS WERE PLACED ON SHIP HOLD TO CONTAIN THE DEFECT. ALL POTENTIALLY AFFECTED PARTS STILL IN GE HEALTHCARE'S POSSESSION HAVE BEEN REWORKED AND RE-INSPECTED PER NEW TEST PROCEDURE. GE HEALTHCARE HAS INITIATED A RECALL OF THE POTENTIALLY DEFECTIVE UNITS UNDER CORRECTION / REMOVAL NUMBER: 2242551-04/29/11-004-R. IT WAS DETERMINED DURING EVAL THAT THE FILTER WAS MANUFACTURED SOMETIME BETWEEN (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PT COULD NOT BE VENTILATED. THE CONNECTION TO THE OXYGEN SOURCE WAS VIA THE ANES CIRCUIT (ANESTHESIA CIRCUIT) CONTAINING AN IN-LINE MOISTURE FILTER (HME FILTER). UPON INSPECTION, THE 5701 HME FILTER WITHIN THE ANESTHESIA CIRCUIT WAS FOUND TO BE OCCLUDED BY A PIECE OF PLASTIC. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANES CIRCUITS ANESTHESIA CIRCUIT CAI VITAL SIGNS, INC. A4U52X1X 11090A

Patients

Seq Age Sex Outcome Treatment
1