FDA Adverse Event Injury Summary report: N

VERSAFITCUP DOUBLE MOBILITY HC LINER

MDR report key: 2096684 · Received May 5, 2011

Report

Report Number
3005180920-2011-00022
Event Type
Injury
Date Received
May 5, 2011
Date of Event
April 12, 2011
Report Date
May 5, 2011
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
MEH
PMA / PMN Number
K092265
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE ACETABULAR LINER WAS JUST RECEIVED AND IT WILL BE ANALYZED. DOCUMENT REVIEW: VERSAFITCUP DM HC LINER REF. (B)(4)/LOT. 091558. ALL PARAMETERS WERE FOUND TO BE CONFORMING TO SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. THE WASHING CYCLES FOR PLASTIC COMPONENTS WERE RUN ACCORDING TO SPECIFICATIONS AND NO ALARM OR DEVIATION OCCURRED DURING OPERATION. THE STERILIZATION CYCLES WERE PERFORMED ACCORDING TO SPECIFICATIONS VALID AT THE TIME OF PRODUCTION. FORTY ONE LINERS BELONGING TO THIS LOT (TOTAL 46 PCS) HAVE BEEN IMPLANTED AND NO INCIDENTS HAVE BEEN REPORTED UP TO NOW. AMISTEM H FEM STEM SIZE 4 - REF. (B)(4)/LOT. 100539. COCR BALL HEAD 28 M - REF. (B)(4)/LOT. 100528. ALL PARAMETERS WERE FOUND TO BE CONFORMING TO SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. THE WASHING CYCLES FOR METAL COMPONENTS WERE RUN ACCORDING TO SPECIFICATIONS AND NO ALARM OR DEVIATION OCCURRED DURING OPERATION. THE STERILIZATION CYCLES WERE PERFORMED ACCORDING TO SPECIFICATIONS VALID AT THE TIME OF PRODUCTION. (B)(4). THE REVISION SURGERY IS ASSOCIATED TO AN INFECTION EVENT. ON THE BASIS OF THE BATCH RECORDS REVIEW AND THE ABSENCE OF INFECTION EVENTS ON THE OTHER PIECES OF THE SAME LOTS, A DEVICE INVOLVEMENT IS HIGHLY UNLIKELY.

Description of Event or Problem · 1

PT HAD LATERAL THIGH PAIN DUE TO MICROMOTION OF THE FEMORAL STEM AS THE PROXIMAL 2/3RDS WERE NOT FIXED. THE PRE-OPERATIVE PLAN WAS TO REVISE THE STEM AND HEAD WITH PROXIMAL FIXATION, BUT DURING SURGERY, THE HC LINER WAS YELLOW DUE TO BACTERIAL INFECTION. REVISED HC LINER AND USED THE EXPLANTED FEMORAL STEM TO CREATE THE ANTIBIOTIC BONE CEMENT SPACER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSAFITCUP DOUBLE MOBILITY HC LINER ACETABULAR HC PE LINER MEH MEDACTA INTERNATIONAL SA 091558

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention COCR BALL HEAD 28 SIZE M: REF. (B)(4)| LOT. 100528| LOT. 100539| AMISTEM H FEM STD SIZE 4: REF. (B)(4)