FDA Adverse Event Malfunction Summary report: N

NEEDLE SPINAL S/SU 22GA 3-1/2IN QUINCKE

MDR report key: 20964749 · Received December 18, 2024

Report

Report Number
2618282-2024-00192
Event Type
Malfunction
Date Received
December 18, 2024
Date of Event
December 9, 2024
Report Date
January 23, 2025
Manufacturer
BECTON DICKINSON
Product Code
BSP
PMA / PMN Number
K210978
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: THREE (3) 22GA X 3.5¿ UNUSED AND SEALED NEEDLES FOR BATCH NUMBER 4123761 WERE RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. THE SAMPLES WERE SUBMITTED FOR VISUAL INSPECTION FIRST. THERE WAS NO POINT DAMAGE, INCLUDING HOOKS, SWAGE MARKS, SCRATCHES, INCOMPLETE OR OPEN POINTS, UNSMOOTH GROUND SURFACES, DISCOLORATION OF ANY KIND OR SILVERS, AND NO PROTRUDERS, NO RECEDERS. THE NEEDLES WERE ALSO SUBMITTED TO TEST FOR ANY CLOGS OR BLOCKAGES. THE SAMPLES COMPLY WITH THE EXPECTED CHARACTERISTICS AND THE REPORTED FAILURE WAS NOT OBSERVED IN THE NEEDLES; THEREFORE, THE REPORTED FAILURE COULD NOT BE VERIFIED. A REVIEW OF THE INTERNAL MANUFACTURING DEVICE RECORDS AND RAW MATERIAL HISTORY FILES FOR THE REPORTED LOT NUMBERS WAS PERFORMED AND NO RECORDED QUALITY PROBLEMS OR REJECTIONS TO THIS INCIDENT WERE FOUND. BASED ON THE QUALITY TEAM'S INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED AS THE INCIDENT COULD NOT BE VERIFIED.

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

MATERIAL# 405181 BATCH# 4123761 VERBATIM: I AM TOLD THAT THESE HAVE BEEN ¿CLOGGING UP¿ WHEN BEING USED. THREE DIFFERENT STAFF MEMBERS HAVE STATED THIS. NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2284418 NEEDLE SPINAL S/SU 22GA 3-1/2IN QUINCKE NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) BSP BECTON DICKINSON 4123761

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown