FDA Adverse Event Injury Summary report: N

AMISTEM-H PROXIMAL COATING FEMORAL

MDR report key: 20961927 · Received December 18, 2024

Report

Report Number
3005180920-2024-01021
Event Type
Injury
Date Received
December 18, 2024
Date of Event
November 18, 2024
Report Date
December 18, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LPH
PMA / PMN Number
K161635
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 20-11-2024: LOT 171686: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23-07-2017. EXPIRATION DATE: 2022-08-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DEPARTMENT: A REVISION SURGERY WAS PERFORMED ABOUT 6.5 YEARS AFTER THE PRIMARY IMPLANTATION OF A THA. THE COMPLAINED OF PAIN AND UNDERWENT FOLLOW-UP X-RAY WHICH REVEALED RADIOLUCENCY LINES AROUND THE STEM, PARTICULARLY IN THE PROXIMAL REGION, ALONG WITH SIGNS OF STRESS SHIELDING AND POSSIBLE SLIGHT SUBSIDENCE. THE BONE AROUND THE DISTAL APEX OF THE STEM EXHIBITED HIGH DENSITY AND CORTICAL THICKENING. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY CEMENTLESS HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS FAILURE CANNOT BE EXACTLY DETERMINED. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DEVICE HISTORY RECORD REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE.

Description of Event or Problem · 0

AT ABOUT 6 YEARS AND 7 MONTHS FROM PRIMARY THE PATIENT CAME IN REPORTING PAIN, DUE TO A POTTED STEM. THE SURGEON REVISED THE STEM, HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1719487 AMISTEM-H PROXIMAL COATING FEMORAL HIP STEM PROXIMAL COATING CEMENTLESS LPH MEDACTA INTERNATIONAL SA 01.18.186 171686

Patients

Seq Age Sex Outcome Treatment
1 60 YR Male Required Intervention